ATAGO PAL-ES3 Digital Hand-held “Pocket” Salt-meter

Category: Digital Salt-meter

The new “Pocket” Salt-meter PAL-ES3 is suitable for inspections carried during the manufacturing process as well as for quality control at food factories. It is the best for the measuring the % concentration of salt in sauces, catsups, seasoning, and soups. The displayed unit is in g/100ml as in the Mohr’s method.

Specifications
Model PAL-ES3
Cat.No. 4233
Scale Salt concentration (g/100ml)
Measurement Method Conductivity method
Measurement Range 0.0 to 3.3% salt concentration
(Note: User must dilute samples to 10% by weight with distilled water before measuring. Unit will automatically multiply readings by 10 to compensate.)
Resolution 0.1g/100ml
Measurement Accuracy Displayed value ±0.6g/100ml Relative precision ±less than 6%
(for measurement value of 10 to 33.0g/100ml)
Temperature compensation range 10 to 40°C (Guaranteed accuracy range 15 to 35°C)
Ambient Temperature 10 to 40°C (Guaranteed accuracy range 15 to 35°C)
Sample Volume 0.3ml or more
Measurement Time 3 seconds
Power Supply Size AAA alkaline
batteries × 2
International Protection Class IP65 Dust-tight and Protected against water jets.
Dimensions & Weight 55(W)×31(D)×109(H)mm, 100g (Main Unit only)
Options

· PAL-CASE : RE-39409

· STRAP : RE-39410

· NaCl Solution 2.50% (±0.05%) AB250 [*1] : RE-120250

[*1] Calibration Certificate : Contact an ATAGO representative for details.

Location

Ask about this product